Total Clinical Chemistry Laboratory Errors and Evaluation of the Analytical Quality Control Using Sigma Metric for
Mulugeta Teshome1, Abebaw Worede2, Daniel Asmelash2
1Department of Medical Laboratory, Dessie Comprehensive Specialized Hospital, Dessie, Ethiopia.
Background:
Currently, the use of clinical laboratory tests is growing at a promising rate and about 80% of the clinical decisions made are based on the laboratory test results. Therefore, it is a major task to achieve quality service. This study was conducted to assess the magnitude of errors in the total testing process of Clinical Chemistry Laboratory and to evaluate analytical quality control using sigma metrics.
Methods:
A cross-sectional study was conducted at Dessie Comprehensive Specialized Hospital Clinical Chemistry Laboratory, Northeast Ethiopia, from 10 February 2020 to 10 June 2020. All Clinical Chemistry Laboratory test requests with their respective samples, external quality control and all daily internal quality control data during the study period were included in the study. Data were collected using a prepared checklist and analyzed using SPSS version 21.
Results:
A total of 4719 blood samples with their test requests were included in the study. Out of 145,383 quality indicators, an error rate of 22,301 (15.3%) was identified in the total testing process. Of the total errors, 76.3% were pre-analytical, 2.1% were analytical and 21.6% were post-analytical errors (p<0.0001). Of the total 14 analytes in the sigma metric evaluation, except ALP, all routine clinical chemistry tests were below the standard (<3). In multivariate logistic regression, the location of patients in the inpatient department was significantly associated with the specimen rejection ((AOR=1.837, 95% CI (1.288-2.618), p=0.001).
Conclusion:
The study found a higher frequency of errors in the total testing process in the Clinical Chemistry Laboratory and almost all test parameters had an unsatisfactory sigma metric value.
Related Concept Videos
Contaminants and Errors
Another key consideration is determining the appropriate number of samples required to...
Systematic Error: Methodological and Sampling Errors
Sampling errors originate from improper sampling methods or the wrong sample population. These errors can be minimized by refining the sampling strategy. Defective instruments or faulty calibrations are the sources of instrumental...
Quality Control
Quality control helps track data, visualize trends, and identify variations, making it easier to detect deviations that may affect the accuracy of an analysis. One way to do this is by generating a quality control chart, which...
Data Validation
Key parameters for method validation include:
Quality Assurance
Random and Systematic Errors