A novel constancy check technique for a well type ionisation chamber using a 90Sr plane-parallel check source
1Radiation Oncology Department, Sir Charles Gairdner Hospital, Hospital Avenue, Nedlands, WA, 6009, Australia. Jonathon.thompson@health.wa.gov.au.
Abstract:
Accurate determination of source strength in brachytherapy is vital to achieving safe and effective radiotherapy treatments. The recommended detectors to be used for determining source strength are well type ionisation chambers and it is essential to the process that they retain a stable response to radiation, thus maintaining a valid calibration. A new Standard Imaging Inc. HDR-1000 Plus well chamber was acquired by GenesisCare Western Australia and commissioned for use clinically. As part of this process, a novel technique was developed to allow constancy checks of the well chamber to be performed using a Sr-90 check source designed for use with plane-parallel chambers (PTW model T48010). The proposed methodology is simple, robust and sufficiently accurate to measure well chamber constancy within ± 0.5%. Most radiotherapy departments will already possess a Sr-90 check source as they are required for external beam dosimetry equipment constancy checks. Therefore, no new additional equipment need be purchased making this method particularly cost-effective.
More Related Videos
07:31Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
06:28Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
Related Concept Videos
Mass Analyzers: Common Types
Atomic Emission Spectroscopy: Lab
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Atomic Absorption Spectroscopy: Instrumentation
The atomizer used in AAS can be either a flame atomizer or an...
Atomic Absorption Spectroscopy: Radiation and Light Sources
Two common narrow-range 'line' sources used in AAS are hollow-cathode lamps (HCLs) and...
Gas Chromatography: Types of Detectors-II