Radiation dosimetry of a clinical prototype dedicated cone-beam breast CT system with offset detector
Hsin Wu Tseng1, Andrew Karellas1, Srinivasan Vedantham1,2
1Department of Medical Imaging, The University of Arizona, Tucson, AZ, USA.
Purpose:
A clinical-prototype, dedicated, cone-beam breast computed tomography (CBBCT) system with offset detector is undergoing clinical evaluation at our institution. This study is to estimate the normalized glandular dose coefficients ( ) that provide air kerma-to-mean glandular dose conversion factors using Monte Carlo simulations.
Materials And Methods:
The clinical prototype CBBCT system uses 49 kV x-ray spectrum with 1.39 mm 1st half-value layer thickness. Monte Carlo simulations (GATE, version 8) were performed with semi-ellipsoidal, homogeneous breasts of various fibroglandular weight fractions ( , chest wall diameters ( cm), and chest wall to nipple length ( ), aligned with the axis of rotation (AOR) located at 65 cm from the focal spot to determine the . Three geometries were considered - -cm detector with no offset that served as reference and corresponds to a clinical CBBCT system, -cm detector with 5 cm offset, and a -cm detector with 10 cm offset.
Results:
For 5 cm lateral offset, the ranged mGy/mGy and reduction in with respect to reference geometry was observed only for 18 cm ( ) and 20 cm ( ) diameter breasts. For the 10 cm lateral offset, the ranged mGy/mGy and reduction in was observed for all breast diameters. The reduction in was , , , , and for 8, 10, 14, 18, and 20 cm diameter breasts, respectively. For a given breast diameter, the reduction in with offset-detector geometries was not dependent on . Numerical fits of were generated for each geometry.
Conclusion:
The and the numerical fit, would be of benefit for current CBBCT systems using the reference geometry and for future generations using offset-detector geometry. There exists a potential for radiation dose reduction with offset-detector geometry, provided the same technique factors as the reference geometry are used, and the image quality is clinically acceptable.