A Low-Power Stable Wideband Current Source for Acupuncture Point Skin Impedance Measurements
Changpei Qiu1, Tianxia Zhao1, Qiuping Li1
1The Key Laboratory of Integrated Microsystems, Peking University Shenzhen Graduate School, Shenzhen 518055, China.
Abstract:
A low-power stable wideband current source for acupuncture point skin impedance measurements has been designed employing a differential architecture and negative feedback. The circuits extend bandwidth to 1 MHz, reducing harmonic distortion to 0.24% at 1 MHz. The output impedance is 37 MΩ at 100 kHz and 11 MΩ at 1 MHz. The stability of the output current of the current source when connected to different loads is below 0.1% at frequencies up to 500 kHz and increases to 0.74% at 1 MHz. The circuit was manufactured in a 0.13-μm CMOS technology and measured results are presented. The area of the current source is 0.09 mm2 and its consumption is 1.2 mW. It is intended for low-power acupuncture point skin impedance measurements.