Reference thresholds for the ER-3A insert earphone
L A Wilber1, B Kruger, M C Killion
1Program in Audiology, Northwestern University, Evanston, Illinois 60208.
The Journal of the Acoustical Society of America
|February 1, 1988
Summary
The ER-3A insert earphone can often replace the TDH-39/MX-41AR without recalibration. Analysis of five studies confirms its reference threshold levels closely match established standards for audiology.
Area of Science:
- Audiology
- Acoustics
- Hearing Instrument Technology
Background:
- The ER-3A insert earphone is increasingly considered a substitute for the TDH-39/MX-41AR.
- Existing data requires analysis to establish reliable reference threshold levels for the ER-3A.
Purpose of the Study:
- To analyze data from multiple studies to estimate reference threshold levels for the ER-3A earphone.
- To compare these estimates with manufacturer specifications and established audiological standards.
Main Methods:
- Meta-analysis of data from five recent studies involving the ER-3A earphone.
- Conversion of sound pressure levels (SPLs) from a 2-cc coupler to Zwislocki-coupler-equivalent SPLs.
- Comparison of results with provisional reference threshold SPLs and revised minimum audible pressures.
Main Results:
- Mean data from the five studies align within 1 dB of the ER-3A manufacturer's provisional reference threshold SPLs (HA-1, 2-cc coupler).
- Converted SPLs show agreement within 1.5 dB with Killion's revised minimum audible pressures across audiometric frequencies.
- The ER-3A earphone demonstrates reliable performance comparable to established standards.
Conclusions:
- The ER-3A insert earphone can be used as a substitute for the TDH-39/MX-41AR without recalibration in most audiological applications.
- Both manufacturer's provisional SPLs and the study's averaged results provide accurate reference levels for the ER-3A.
- This analysis provides a validated basis for using the ER-3A in hearing assessments.