Choice of Continuous Glucose Monitoring Systems May Affect Metrics: Clinically Relevant Differences in Times in
Guido Freckmann1, Stefan Pleus1, Sebastian Schauer1
1Institut für Diabetes-Technologie, Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm, Ulm, Germany.
Background:
Continuous glucose monitoring-derived parameters are becoming increasingly important in the treatment of people with diabetes. The aim of this study was to assess whether these parameters, as calculated from different continuous glucose monitoring systems worn in parallel, are comparable. In addition, clinical relevance of differences was investigated.
Methods:
A total of 24 subjects wore a FreeStyle Libre (A) and a Dexcom G5 (B) sensor in parallel for 7 days. Mean glucose, coefficient of variation, glucose management indicator and time spent in different glucose ranges were calculated for each system. Pairwise differences between the two different continuous glucose monitoring systems were computed for these metrics.
Results:
On average, the two CGM systems indicated an identical time in range (67.9±10.2 vs. 67.9±11.5%) and a similar coefficient of variation; both categorized as unstable (38.1±5.9 vs. 36.0±4.8%). In contrast, the mean time spent below and above range, as well as the individual times spent below, in and above range differed substantially. System A indicated about twice the time spent below range than system B (7.7±7.2 vs. 3.8±2.7%, p=0.003). This could have led to different therapy recommendations in approximately half of the subjects.
Discussion:
The differences in metrics found between the two continuous glucose monitoring systems may result in different therapy recommendations. In order to make adequate clinical decisions, measurement performance of CGM systems should be standardized and all available information, including the HbA1c, should be utilized.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Oral Hypoglycemic Agents: Glinides
Glucose Homeostasis: Regulation of Blood Glucose
During fasting, when blood glucose levels are low, the pancreas secretes glucagon. it...
Therapeutic Drug Monitoring: Overview and Classification
Hypoglycemia and Glucagon
Diabetes Mellitus: Type 2 and Gestational