M Jones1, E E Eidbo, E R Rodriguez
1Surgery Branch, National Heart, Lung, and Blood Institute, Bethesda, MD 20892.
Abstract:
Damage to the posterolateral wall of the left ventricle was found in eight of approximately 700 sheep undergoing mitral valvular replacement as part of animal model studies of bioprosthetic valves. The damage consisted of left ventricular aneurysms in five animals, subacute rupture of the left ventricle in one, acute left ventricular laceration in one, and endocardial scarring in one. Six of the eight bioprostheses were bovine pericardial valves, including five low-profile valves and one standard valve; of the two porcine bioprostheses, one was intentionally oversized and the other was a low-profile supra-annular valve. In each of these animals the damage appeared to have been caused by contact between the most posterior strut of the bioprosthesis and the left ventricular wall.
Related Concept Videos
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...