Purpose-Built Transcatheter Cavopulmonary Anastomosis Device Requirements: Multi-Modality Imaging Evaluation
Hannah R El-Sabrout1, Justin R Ryan2, Sanjeet R Hegde1
1Division of Pediatric Cardiology, Rady Children's Hospital and UC San Diego School of Medicine, 3020 Children's Way, San Diego, CA, USA.
A less-invasive transcatheter approach for single ventricle patients may reduce open-heart surgeries. Ideal device measurements for a superior cavopulmonary anastomosis (or bidirectional Glenn shunt) are 10 mm diameter and 20-25 mm length.
Area of Science:
- Pediatric cardiology
- Medical imaging
- Surgical innovation
Background:
- Patients with functional single ventricle often require multiple palliative open-heart surgeries, including superior cavopulmonary anastomosis (or bidirectional Glenn shunt).
- A less-invasive transcatheter approach could potentially decrease patient morbidity associated with traditional surgical methods.
Purpose of the Study:
- To evaluate imaging modalities for pre-operative assessment of patients undergoing Glenn shunt procedures.
- To determine optimal device dimensions for a transcatheter superior cavopulmonary anastomosis.
Main Methods:
- Analysis of pre-Glenn X-ray contrast angiography (XA), cardiac computed tomography (CT), and cardiac magnetic resonance (CMR) studies from 139 patients.
- Measurement of superior vena cava (SVC) and right pulmonary artery (RPA) dimensions, SVC-RPA angle, and potential device lengths across imaging modalities.
- Assessment of image quality for each modality.
Main Results:
- Mean SVC and RPA measurements were comparable across XA, CT, and CMR.
- Ideal device length varied by imaging modality, with CMR suggesting longer devices.
- Image quality was sufficient for most CT and XA studies, and for most SVC and RPA measurements in MRI.
Conclusions:
- Parametric modeling suggests a 10 mm diameter and 20-25 mm length device is ideal for transcatheter superior cavopulmonary anastomosis.
- Imaging modalities like CT and CMR provide valuable pre-operative data for planning less-invasive procedures.