Kenneth I Berger1, Ori Adam2, Roberto Walter Dal Negro3
1Division of Pulmonary Critical Care and Sleep Medicine, NYU Grossman School of Medicine, Bellevue Hospital, New York, NY; André Cournand Pulmonary Physiology Laboratory, Bellevue Hospital, New York, NY.
Background:
Current techniques for measuring absolute lung volumes rely on bulky and expensive equipment and are complicated to use for the operator and the patient. A novel method for measurement of absolute lung volumes, the MiniBox method, is presented.
Research Question:
Across a population of patients and healthy participants, do values for total lung capacity (TLC) determined by the novel compact device (MiniBox, PulmOne Advanced Medical Devices, Ltd.) compare favorably with measurements determined by traditional whole body plethysmography?
Study Design And Methods:
A total of 266 participants (130 men) and respiratory patients were recruited from five global centers (three in Europe and two in the United States). The study population comprised individuals with obstructive (n = 197) and restrictive (n = 33) disorders as well as healthy participants (n = 36). TLC measured by conventional plethysmography (TLCPleth) was compared with TLC measured by the MiniBox (TLCMB).
Results:
TLC values ranged between 2.7 and 10.9 L. The normalized root mean square difference (NSD) between TLCPleth and TLCMB was 7.0% in healthy participants. In obstructed patients, the NSD was 7.9% in mild obstruction and 9.1% in severe obstruction. In restricted patients, the NSD was 7.8% in mild restriction and 13.9% in moderate and severe restriction. No significant differences were found between TLC values obtained by the two measurement techniques. Also no significant differences were found in results obtained among the five centers.
Interpretation:
TLC as measured by the novel MiniBox system is not significantly different from TLC measured by conventional whole body plethysmography, thus validating the MiniBox method as a reliable method to measure absolute lung volumes.
Related Concept Videos
Lung Capacity
Respiratory Volumes and Capacities I
Respiratory Volumes and Capacities
Respiratory Volumes
Tidal Volume (TV) Tidal volume (TV) is the air inhaled or exhaled in a...
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Pulmonary Function Tests
Pulmonary Function Tests are crucial diagnostic tools for assessing respiratory function, particularly in patients with chronic respiratory disorders. They comprehensively evaluate lung volumes, ventilatory function, breathing mechanics, diffusion, and gas exchange. These tests help diagnose pulmonary diseases and play a significant role in monitoring disease progression, evaluating disability, and assessing response to therapy.
PFTs involve using a spirometer, a...