Assessment of Plasma Coagulation on Liver Tissue in a Large Animal Model In Vivo
Published on: August 4, 2018
Argon Plasma Coagulation: Elucidation of the Mechanism of Gas Embolism
Erik E Folch1, Catherine L Oberg2, Atul C Mehta3
1Division of Pulmonary and Critical Care Medicine, Section of Interventional Pulmonary, Massachusetts General Hospital, Harvard Medical School, Boston, Massachusetts, USA.
Background:
Argon plasma coagulation (APC) is a tool used in the management of tracheobronchial obstruction or bleeding. Complications include gas embolism which can cause devastating effects including hemodynamic instability, cardiac arrest, and stroke. Multiple theories as to how gas embolism occurs with APC have been postulated; however, none have identified the exact mechanism.
Objectives:
To identify the mechanism by which APC causes gas embolism in the tracheobronchial tree.
Methods:
Using an explanted porcine tracheobronchial tree with lung parenchyma, the APC catheter was applied through noncontact and direct contact to the endobronchial airway mucosa via flexible bronchoscopy. This was done at multiple gas flow settings and pulse durations. Visual changes in the mucosa were photographed, videoed, and described.
Results:
Gross evidence of submucosal gas transfer occurred when the APC catheter was in direct contact with the mucosa at all gas flow settings in all applications, despite using shorter pulse durations. Whenever the catheter was not in contact with the mucosa, there was no transfer of gas at any gas flow setting or pulse duration.
Conclusions:
Direct mucosal contact with the APC probe leads to submucosal gas deposition and is a likely mechanism for gas entry into the intravascular space. In reported cases of APC-associated gas embolism, presence of a vascularized endobronchial tumor may have increased the risk of gas tracking into the intravascular space. Care should be taken when applying APC during brisk bleeding or limited vision, as inadvertent mucosal contact may occur and could increase the risk of gas embolism.
Related Concept Videos
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Atomic Emission Spectroscopy: Overview
Pulmonary Embolism I: Introduction
Noble Gases
The elements in group 18 are noble gases (helium, neon, argon, krypton, xenon, and radon). They earned the name “noble” because they were assumed to be nonreactive since they have filled valence shells. In 1962, Dr. Neil Bartlett at the University of British Columbia proved this assumption to be false.
Physical Principles Governing Gas Exchange
Gas Laws Governing Respiration
The behavior of gases is guided by Dalton's Law of partial pressures and Henry's Law.
Dalton's Law asserts that the total...
Coagulation
During the coagulation phase, clotting factors, or procoagulants, play a vital role in initiating and progressing the coagulation cascade. This cascade is a series of reactions...