UriSed 3 PRO automated microscope in screening bacteriuria at region-wide laboratory organization
Timo Kouri1, Tanja Holma2, Vesa Kirjavainen2
1Department of Clinical Chemistry, HUSLAB, Helsinki University Hospital, HUS Diagnostic Center, FIN-00029 HUS, Helsinki, Finland; Department of Clinical Chemistry, University of Helsinki, Finland.
Background And Aims:
We assessed the possibility to rule out negative urine cultures by counting with UriSed 3 PRO (77 Elektronika, Hungary) at Helsinki and Uusimaa Hospital District.
Materials And Methods:
Bacteria counting of the UriSed 3 PRO automated microscope was verified with reference phase contrast microscopy against growth in culture. After acceptance into routine, results of bacteria and leukocyte counting from 56 426 specimens with eight UriSed 3 PRO instruments were compared against results from parallel samples cultured on chromogenic agar. Laboratory data including preanalytical details were accessed through the regional database of the Helsinki and Uusimaa Hospital District.
Results:
A combined sensitivity of 87-92% and a negative predictive value of 90-96% with a specificity of 54-50% was reached, depending on criteria. Preanalytical data (incubation time in bladder) combined with the way of urine collection would improve these figures if reliable.
Conclusions:
Complex patient populations, regional logistics and data interfases, and economics related to increased costs of additional particle counts against costs of screening cultures of all samples, did not support adaptation of a screening process of urine cultures. This conclusion was made locally, and may not be valid elsewhere.