Evaluation of the ARKRAY HA-8190V instrument for HbA1c
Eline A E van der Hagen1,2, Sanne Leppink2, Karin Bokkers2
1European Reference Laboratory for Glycohemoglobin, Queen Beatrix Hospital, Winterswijk, The Netherlands.
Objectives:
Hemoglobin A1c (HbA1c) is a valuable parameter in the monitoring of diabetic patients and increasingly in diagnosis of diabetes. Manufacturers continuously optimize instruments, currently the main focus is to achieve faster turnaround times. It is important that performance specifications remain of high enough standard, which is evaluated in this study for the new ARKRAY HA-8190V instrument.
Methods:
The Clinical and Laboratory Standards Institute (CLSI) protocols EP-5, EP-9 and EP-10 were applied to investigate imprecision, bias and linearity. In addition potential interferences, performance in External Quality Assessment (EQA) and performance against the HA-8180V instrument in 220 clinical samples was evaluated.
Results:
The HA-8190V demonstrates a CV of ≤0.8% in IFCC SI units (≤0.6% National Glycohemoglobin Standardization Program [NGSP]) at 34 and 102 mmol/mol levels (5.3 and 11.5% NGSP) and a bias of -0.1 mmol/mol (-0.01% NGSP) at a concentration of 50 mmol/mol (6.7% NGSP), but with a significant slope as compared to target values. This results in a bias of -1.0 and 0.9 mmol/mol (-2.0 and 0.9% NGSP) at the 30 and 70 mmol/mol (4.9 and 8.6% NGSP) concentration level. Simulation of participation in the IFCC certification programme results in a Silver score (bias -0.1 mmol/mol, CV 1.1%). Interference in the presence of the most important Hb variants (AS, AC, AE, AD) and elevated HbA2 and HbF concentrations is less than 3 mmol/mol (0.3% NGSP) at a concentration of 50 mmol/mol (6.7% NGSP).
Conclusions:
Analytical performance of the HA-8190V is very good, especially with respect to precision and HbA1c quantification in the presence of the most common Hb variants.
Related Concept Videos
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Measurement of Blood Pressure
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...