Related Experiment Video
Updated: Nov 18, 2025

Assessment and Evaluation of the High Risk Neonate: The NICU Network Neurobehavioral Scale
Published on: August 25, 2014
The validity of the Language Environment Analysis system in two neonatal intensive care units
Eva Ståhlberg-Forsén1, Anette Aija2,3, Birgit Kaasik2
1University of Helsinki, Helsinki, Finland.
Aim:
To evaluate the validity of the Language Environment Analysis (LENA) system's automatic measures in two neonatal intensive care units supporting parent-infant closeness, and in two Finno-Ugric languages: Finnish and Estonian.
Methods:
The sound environment of 70 very preterm infants was recorded for 16 h in the neonatal intensive care units with the LENA system roughly at the gestational age of 32 (+2) weeks. Of these, the recordings of 14 infants (20%, two 5-min samples with a high percentage of speech, totally 140 min) were analysed in detail and in two different ways. Parental closeness diaries were used to document the presence of the parents. Agreements between LENA system and human coder estimates were analysed.
Results:
Findings showed a high variation in agreements. The highest agreements were found in female and adult word counts (r = 0.91 and 0.95). The agreements for child vocalisation count, conversational turns and silence were modest or low (r = -0.03 to 0.64).
Conclusion:
Our study provides novel information on the validity of the LENA system in the neonatal intensive care unit. Findings show that the LENA system provides valid information on adult words, but LENA estimates for child vocalisations were less valid at this early age.
More Related Videos
11:14A Novel Experimental and Analytical Approach to the Multimodal Neural Decoding of Intent During Social Interaction in Freely-behaving Human Infants
Published on: October 4, 2015
07:31Defining the Role Of Language in Infants' Object Categorization with Eye-tracking Paradigms
Published on: February 8, 2019