Melted delivery hose--a complication of a heated humidifier
1Department of Anaesthesia, Faculty of Medicine, University of British Columbia, Vancouver.
Abstract:
The polyvinylchloride delivery hose of a Fisher & Paykel dual servo heated anaesthetic humidifier was melted by the intraluminal hose heater, causing a major leakage of anaesthetic gases during a surgical procedure. Under experimental conditions, melting could be reproduced using polyvinylchloride but not polyester elastomer tubings. Only tubings made of materials with high melting point should be used as delivery hose in heated humidifiers with hose heaters.
Related Concept Videos
Phase Transitions: Vaporization and Condensation
Heating and Cooling Curves
For instance, the addition of heat raises the temperature of a solid; the amount of heat absorbed depends on the heat capacity of the solid (q = mcsolidΔT). According to thermochemistry, the relation between the amount of heat absorbed or released by a substance, q, and its...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Masonry in Cold and Hot Weather Conditions
Other key practices include keeping masonry units and sand dry and...
Mortar Joint Deterioration in Masonry
The...