Related Experiment Video
Updated: Nov 18, 2025

05:14
Comparison of Agreement and Accuracy using Binocular Wavefront Optometer with Autorefractor and Phoropter
Published on: September 16, 2025
343
Accuracy of partial coherence interferometry in patients with large inter-eye axial length difference
Christine A Petersen1, Daniel C Terveen2, Tyler Quist2
1Department of Ophthalmology, University of Washington, Seattle, Washington, United States of America.
Plos One
|February 9, 2021
Summary
Partial coherence interferometry (PCI) is accurate for cataract surgery patients with large inter-eye axial eye length (AEL) differences. This finding suggests confirmatory measurements may not be needed, streamlining pre-operative assessments.
Area of Science:
- Ophthalmology
- Biomedical Optics
- Refractive Surgery
Background:
- Cataract surgery requires precise biometry for accurate intraocular lens (IOL) power calculation.
- Inter-eye differences in axial eye length (AEL) can pose challenges for refractive outcomes.
- Partial coherence interferometry (PCI) is a common biometry technique.
Purpose of the Study:
- To evaluate the accuracy of PCI in patients with significant inter-eye AEL differences (> 0.30 mm).
- To compare refractive prediction errors between patients with and without large inter-eye AEL differences.
Main Methods:
- Retrospective review of 2212 eyes from 1617 patients undergoing cataract surgery.
- Inclusion criteria: inter-eye AEL difference > 0.30 mm (cases) vs. no significant difference (controls).
- IOL power calculated using SRK/T, Holladay II, and Hoffer Q formulae; refractive prediction error analyzed.
Main Results:
- 131 case eyes and 131 control eyes were matched.
- Mean absolute error (MAE) ranged from 0.47 D to 0.69 D and was not statistically different between groups.
- No significant difference in refractive prediction error or proportion of eyes with >= 0.50 D or 1.00 D error.
Conclusions:
- Good-quality PCI data demonstrates equal accuracy in eyes with or without large inter-eye AEL differences.
- Confirmatory AEL measurements may be unnecessary for patients with substantial inter-eye AEL disparities.