Coronary compression testing by balloon interrogation during pulmonary valve implantation: room for doubt?
Gareth J Morgan1,2,3, Salvador A Rodriguez1,2, Ryan Leahy1,2
1The Heart Institute, Children's Hospital Colorado, Aurora, CO, USA.
The balloon coronary test is unreliable for predicting coronary compression during pulmonary valve implantation. This study found a significant rate of inaccurate test results, questioning its clinical utility.
Area of Science:
- Cardiology
- Interventional Cardiology
- Medical Imaging
Background:
- The balloon coronary test is widely used to assess coronary compression risk during pulmonary valve implantation.
- However, cases of coronary compression have occurred despite low-risk balloon test results, raising concerns about its accuracy.
Purpose of the Study:
- To evaluate the reliability of the balloon coronary test in predicting coronary compression during percutaneous pulmonary valve implantation.
- To identify factors influencing the accuracy of this test.
Main Methods:
- Retrospective review of 36 patients who underwent pulmonary valve implantation with suitable imaging for quantitative analysis.
- Analysis focused on spatial discrepancy between the testing balloon position and the final implanted valve position.
Main Results:
- The balloon coronary test was found to be inaccurate in 36.1% of cases.
- Right ventricular outflow tract substrate was a significant predictor of test accuracy.
- Testing balloon type and implanted valve size were associated with test accuracy.
Conclusions:
- Balloon coronary testing is not a reliable predictor of final valve position relative to coronary arteries.
- The test may lead to a high false positive rate for coronary compression risk in pulmonary valve implantation.