Preclinical Evaluation of the Stealth Autoguide Robotic Guidance Device for Stereotactic Cranial Surgery: A Human Cadaveric Study
Related Experiment Video
Updated: Nov 18, 2025

Operative Technique and Nuances for the Stereoelectroencephalographic SEEG Methodology Utilizing a Robotic Stereotactic Guidance System
Published on: June 9, 2023
David Brandman1,2, Murray Hong2, David B Clarke3,4
1Department of Surgery (Neurosurgery), Dalhousie University, Halifax, Nova Scotia, Canada.
Introduction:
Stereotactic procedures are routinely performed for brain biopsies, deep brain stimulation, and placement of stereoelectroencephalography (SEEG) electrodes for epilepsy. The recently developed Stealth Autoguide (Medtronic, Minneapolis, MN, USA) device does not require patients to don a stereotactic frame. In this preclinical study, we sought to quantitatively compare the Stealth Autoguide robotic system to 2 devices commonly used in clinical practice: the Navigus biopsy system (Medtronic) and the Leksell stereotactic frame (Elekta Ltd., Stockholm, Sweden).
Methods:
In the first experimental setup, we compared target accuracy of the Stealth Autoguide to the Navigus system by using phantom heads filled with gelatin to simulate the brain tissue. In the second experimental setup, we inserted SEEG electrodes to targets within cadaveric heads in a simulated operating room environment.
Results:
Using a homogeneous gelatin-filled phantom 3D reconstruction of a human head, we found that using the Stealth Autoguide system, while maintaining accuracy, was faster to use than the Navigus system. In our simulated operating room environment using nonliving human cadaveric heads, we found the accuracy of the Stealth Autoguide robotic device to be comparable to that of the Leksell frame.
Discussion/Conclusion:
These results compare the use of the Stealth Autoguide robotic guidance system with commonly used stereotactic devices, and this is the first study to compare its use and accuracy with the Leksell frame. These findings provide mounting evidence that Stealth Autoguide will have potential clinical uses in various stereotactic neurosurgical procedures.
More Related Videos
11:28Robotic-Guided Stereoelectroencephalography for Invasive Epilepsy Monitoring
Published on: June 13, 2025
09:00Investigating the Function of Deep Cortical and Subcortical Structures Using Stereotactic Electroencephalography: Lessons from the Anterior Cingulate Cortex
Published on: April 15, 2015