Conversion factor for size specific dose estimation of head CT scans based on age, for individuals from 0 up to 18
Edilaine Honorio da Silva1, Oswaldo Baffa1, Jorge Elias2
1Department of Physics, FFCLRP, University of São Paulo, Ribeirão Preto, Brazil.
Abstract:
Assessing the radiation doses received by patients in computed tomography is still challenging. To overcome this, the American Association of Physicists in Medicine has introduced the concept of the size specific dose estimate (SSDE). However, the calculation of SSDE for head CT scans requires the knowledge of attenuation characteristics of the volume scanned, making its implementation in the daily clinical workflow cumbersome. In this study, we defined conversion coefficients from CTDIvol,16cmto SSDE for head CT scans based solely on the age of the patient. Using the head circumference-for-age from the child growth standards of the World Health Organization (WHO), the effective diameter-for-age was calculated for male and female individuals from 0 to 60 months-old. The effective diameter was converted into a water equivalent diameter-for-age, using a correlation established from the measurements of both quantities in 295 exams of male and female patients, from 0 to 18 years-old. WHO-estimated water equivalent diameter-for-age was validated against the measured water equivalent diameter-for-age. The head circumference-for-age from WHO was extrapolated for male and females individuals up to 18 years-old and their respective water equivalent diameter were estimated. Finally, the SSDE was calculated for all the CT head scans performed in a 9-years period in patients aged from 0 to 18 years old. Typical values of CTDIvol,16cmand DLP were also defined. SSDE varied from 0.80 up to 1.16 of the CTDIvol,16cm, depending on sex and age of the patient. WHO-estimated water equivalent diameter-for-age differed less than 20% from the measured water equivalent diameter-for-age. Typical values of SSDE varied from 28.5 up to 38.9 mGy, while typical values ranged from 30.9 up to 47.6 mGy for the CTDIvol,16cmand from 417.6 up to 861.1 mGy*cm for the DLP. SSDE can be directly calculated for head CT scans once the age of the patient is known.
More Related Videos
Related Concept Videos
Biological Effects of Radiation
Drug Dosing: Infants and Children
Dose Size and Dosing Frequency: Determination Methods
Drug Dosing in Renal Diseases: Estimation of Glomerular Filtration Rate Based on Serum Creatinine Concentration
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT