A Comparative Study Between the Goldmann Applanation Tonometer and the Non-Contact Air-Puff Tonometer (Huvitz HNT
Ricardo Alexandre Stock1, Carine Ströher1, Rodrigo Rosa Sampaio1
1Medical School of the University of the West of Santa Catarina (Universidade do Oeste de Santa Catarina - UNOESC), Joaçaba, Santa Catarina, Brazil.
Clinical Ophthalmology (Auckland, N.Z.)
|February 12, 2021
Summary
The non-contact air-puff tonometer (NCT) may overestimate intraocular pressure (IOP) compared to the Goldmann applanation tonometer (GAT), especially in higher IOP ranges. While comparable in lower ranges, NCT shows significant differences in moderate to high IOP measurements.
Area of Science:
- Ophthalmology
- Medical Devices
- Diagnostic Technology
Background:
- Intraocular pressure (IOP) measurement is crucial for diagnosing and managing glaucoma.
- The Goldmann applanation tonometer (GAT) is the gold standard for IOP measurement.
- Non-contact air-puff tonometers (NCT) offer a potentially faster and more convenient alternative.
Purpose of the Study:
- To evaluate the non-contact air-puff tonometer (NCT) as a viable alternative to the Goldmann applanation tonometer (GAT) for intraocular pressure (IOP) assessment.
- To compare the accuracy and reliability of IOP measurements obtained from NCT and GAT across different pressure ranges and age groups.
Main Methods:
- An observational, cross-sectional, and comparative study was conducted.
- Intraocular pressure was measured in 180 eyes from 90 patients using both GAT and NCT.
- Data were analyzed quantitatively, comparing measurements between the two devices and across defined IOP and age groups.
Main Results:
- Overall mean IOP measurements differed significantly between NCT (14.12 mmHg) and GAT (12.98 mmHg) (p=0.0018).
- NCT measurements were comparable to GAT in the lower IOP range (10-15 mmHg, p=0.3100) but significantly overestimated IOP in higher ranges (16-19 mmHg and ≥20 mmHg).
- Significant differences were also observed in older age groups (40-59 and ≥60 years), with NCT consistently yielding higher readings.
Conclusions:
- The non-contact air-puff tonometer (NCT) demonstrates potential for IOP measurement but shows overestimation in moderate to high intraocular pressure ranges compared to the Goldmann applanation tonometer (GAT).
- While NCT may be suitable for screening in lower IOP ranges, GAT remains essential for accurate assessment, particularly when elevated IOP is suspected or confirmed.