Related Experiment Video
Updated: Nov 17, 2025

Method and Instrumented Fixture for Femoral Fracture Testing in a Sideways Fall-on-the-Hip Position
Published on: August 17, 2017
Decreased stress shielding with a PEEK femoral total knee prosthesis measured in validated computational models
Lennert de Ruiter1, Kathryn Rankin2, Martin Browne2
1Orthopaedic Research Lab, Radboud University Medical Center Nijmegen, the Netherlands.
Abstract:
Due to their high stiffness, metal femoral implants in total knee arthroplasty may cause stress shielding of the peri-prosthetic bone, which can lead to loss of bone stock. Using a polymer (PEEK) femoral implant reduces the stiffness mismatch between implant and bone, and therefore has the potential to decrease strain shielding. The goal of the current study was to evaluate this potential benefit of PEEK femoral components in cadaveric experiments. Cadaveric femurs were loaded in a materials testing device, while a 3-D digital image correlation set-up captured strains on the surface of the intact femurs and femurs implanted with PEEK and CoCr components. These experimental results were used to validate specimen-specific finite element models, which subsequently were used to assess the effect of metal and PEEK femoral components on the bone strain energy density. The finite element models showed strain maps that were highly comparable to the experimental measurements. The PEEK implant increased strain energy density, relative to the preoperative bone and compared to CoCr. This was most pronounced in the regions directly under the implant and near load contact sites. These data confirm the hypothesis that a PEEK femoral implant can reduce peri-prosthetic stress shielding.

