The Use of Mixed Reality in Custom-Made Revision Hip Arthroplasty: A First Case Report
Published on: August 4, 2022
Can a simple iPad app improve C-arm based component position in anterior THA?
Ulrich Bechler1,2, Bernhard Springer1, Kilian Rueckl1
1Hospital for Special Surgery, 535 East 70th Street, New York, NY, 10021, USA.
A new software app accurately measures acetabular component position and leg length during direct anterior total hip arthroplasty (THA). This technique offers a simple and reliable method for improving surgical outcomes in THA procedures.
Area of Science:
- Orthopedic Surgery
- Medical Imaging Technology
- Biomechanical Engineering
Background:
- Accurate acetabular component positioning is crucial for optimal hip biomechanics and reducing dislocation risk in total hip arthroplasty (THA).
- C-arm imaging is a cost-effective tool for THA but can present parallax challenges for evaluating component placement.
- Direct anterior THA offers potential benefits, necessitating precise intraoperative control of component positioning and leg length.
Purpose of the Study:
- To evaluate the efficacy of a software app-based measurement technique for controlling acetabular component position and leg length during direct anterior THA.
- To assess the accuracy of intraoperative measurements obtained via the software app compared to standard postoperative radiographs.
Main Methods:
- Prospective study involving 93 patients undergoing direct anterior THA.
- Utilized C-arm imaging and an iPad software app for intraoperative measurement of acetabular anteversion, inclination, and leg length discrepancy.
- Compared intraoperative app measurements with standard postoperative AP-pelvis radiographs.
Main Results:
- Strong correlations were found between intraoperative and postoperative measurements for anteversion (r=0.701), inclination (r=0.816), and leg length discrepancy (r=0.542).
- Mean intraoperative anteversion was 20.5° vs. 20.9° postoperatively; inclination was 40.5° vs. 40.7°.
- Mean intraoperative leg length discrepancy was 0.9 mm vs. 0.6 mm postoperatively, demonstrating high accuracy.
Conclusions:
- The evaluated software app provides a simple and accurate method for measuring intraoperative acetabular cup position in direct anterior THA.
- This technology aids in achieving precise leg length equalization during the procedure.
- The app-based approach enhances intraoperative decision-making and can contribute to improved patient outcomes.