Related Experiment Video
Updated: Nov 17, 2025

Voluntary Breath-hold Technique for Reducing Heart Dose in Left Breast Radiotherapy
Published on: July 3, 2014
Do low dose CT-KUBs really expose patients to more radiation than plain abdominal radiographs?
Bob Yang1, Noorunisa Suhail1, Johan Marais1
1Salisbury District Hospital, Salisbury, UK.
Background:
Urolithiasis patients often require frequent urinary tract imaging, leading to high radiation exposure. CT Kidney-Ureter-Bladder (CT-KUB) is the gold standard in urolithiasis detection, however it is thought to harbour significant radiation load. Urologists have therefore utilised abdominal radiographs (XR-KUB) as an alternative, though with markedly lower sensitivity and specificity. We present the first contemporary UK study comparing the effective doses of XR-KUBs with low dose CT-KUBs.
Method:
Fifty-three patients were retrospectively identified in a single centre who underwent both a XR-KUB and a CT-KUB in 2018. Effective-Dose was measured by converting the recorded 'Dose Area/Length Product' via the International Commission on Radiological Protection formula.
Results:
The average effective dose of XR-KUBs and low dose CT-KUBs were 5.10 mSv and 5.31 mSv respectively. Thirty-four percent (18/53) of patients had a XR-KUBs with a higher effective dose than their low dose CT-KUB. Patients with higher Weight, BMI and AP diameter had higher effective doses for both their XR and low dose CT-KUBs. All patients in our study weighing over 92 kg or with a BMI greater than 32 had a XR-KUBs with a higher effective dose than their low dose CT-KUB.
Conclusion:
This data supports moving away from XR-KUBs for the investigation of urolithiasis, particularly in patients with a high BMI.
Related Concept Videos
Radiological Investigation I: X-ray and CT
Imaging Studies III: Computed Tomography
Imaging Studies I: Kidney, Ureter, and Bladder Studies
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
X-ray Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...