Systematic Variability in ICG Recordings Results in ICG Complex Subtypes - Steps Towards the Enhancement of ICG
Sara Benouar1,2, Abdelakram Hafid1,2, Mokhtar Attari1
1Laboratory of Instrumentation, University of Sciences and Technology Houari Boumediene, Algiers, Algeria.
Abstract:
The quality of an impedance cardiography (ICG) signal critically impacts the calculation of hemodynamic parameters. These calculations depend solely on the identification of ICG characteristic points on the ABEXYOZ complex. Unfortunately, contrary to the relatively constant morphology of the PQRST complex in electrocardiography, the waveform morphology of ICG data is far from stationary, which causes difficulties in the accuracy of the automated detection of characteristic ICG points. This study evaluated ICG recordings obtained from 10 volunteers. The results indicate that there are several different waveforms for the ABEXYOZ complex; there are up to five clearly distinct waveforms for the ABEXYOZ complex in addition to those that are typically reported. The differences between waveform types increased the difficulty of detecting ICG points. To accurately detect all ICG points, the ABEXYOZ complex should be analyzed according to the corresponding waveform type.
More Related Videos
09:34A Combinatorial Single-cell Approach to Characterize the Molecular and Immunophenotypic Heterogeneity of Human Stem and Progenitor Populations
Published on: October 25, 2018
06:51Confocal Microscopy Reveals Cell Surface Receptor Aggregation Through Image Correlation Spectroscopy
Published on: August 2, 2018
Related Concept Videos
Immunocytochemistry and Immunohistochemistry
These...
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...