Related Experiment Video
Updated: Nov 17, 2025

Myocardial Infarction by Percutaneous Embolization Coil Deployment in a Swine Model
Published on: November 4, 2021
Periprocedural safety of saccular aneurysm embolization with the Penumbra SMART Coil System: a SMART registry subset
Robert M Starke1, Min S Park2, Richard Bellon3
1Neurological Surgery, University of Miami Hospital, Miami, Florida, USA.
Background:
Using data from the SMART registry, we report on periprocedural safety of the Penumbra SMART Coil System for endovascular coil embolization of saccular intracranial aneurysms.
Methods:
The SMART registry was a prospective, multi-center registry of site standard of care endovascular coiling procedures performed using at least 75% Penumbra SMART Coil, PC400, and/or POD coils. This subset analysis reports on the periprocedural safety outcomes of the saccular intracranial aneurysm cohort. Predictors of rupture/re-rupture or perforation (RRP), thromboembolic complications, and device- or procedure-related adverse events (AEs) were determined in univariate and multivariate analysis.
Results:
Between June 2016 and August 2018, 851 saccular aneurysm patients (31.0%, 264/851 ruptured) were enrolled across 66 North American centers. Clinically significant (ie, a serious adverse event) RRP occurred in 2.0% (17/851) of cases - 1.9% (5/264) for the ruptured cohort and 2.0% (12/587) for the un-ruptured cohort. Clinically significant thromboembolic events occurred in 3.1% (26/851) of cases - 5.3% (14/264) for the ruptured cohort and 2.0% (12/587) for the un-ruptured cohort. Multivariate predictors of periprocedural RRP were increased packing density and adjunctive treatment with a balloon. For periprocedural thromboembolic events, multivariate predictors were bifurcation location and ruptured status. For device- or procedure-related AEs, multivariate predictors were bifurcation location and adjunctive treatment with stent or balloon.
Conclusion:
The low rates of thromboembolic complications and RRP events demonstrate the adequate safety profile of the SMART Coil System to treat cerebral aneurysms in routine clinical practice.
Trial Registration Number:
NCT02729740.