Related Experiment Video
Updated: Oct 1, 2026

Dynamic Lung Tumor Tracking for Stereotactic Ablative Body Radiation Therapy
Published on: June 7, 2015
An intercomparison of two dynamic treatment techniques, ring scan and spot scan, for head and neck tumors with the
M Takai1, H Blattmann, E Pedroni
1Swiss Institute for Nuclear Research, Villigen.
Abstract:
An evaluation of the ring scan and the spot scan was made for the pion irradiation of head and neck tumors with the Piotron. For the geometry of the Piotron, with its 60 radially converging beams, two scanning techniques have been developed, ring scan and spot scan. They have different characteristics concerning achievable dose distributions and sensitivity to tissue inhomogenities. The optimized 3-dimensional dose distributions for the treatment with ring scan and spot scan techniques were calculated for two examples of the target volume. The comparison of the dose distributions has shown that the ring scan is better in sparing normal tissues than the spot scan for a simple shape target volume but not for an irregular shape target volume with the present status of the technique. The irradiation time needed for the ring scan is longer, for the present examples three times, than for the spot scan. From the practical view point the spot scan is preferable to the ring scan for the treatment of head and neck tumors with the Piotron.