Ulnohumeral joint static cartilage compression is affected by radial head implant size
John Elfar1, Michael Caldwell2, Robert Mason3
1The Pennsylvania State University, Center for Orthopaedic Research and Translational Science, 500 University Drive, H089 Hershey, PA 17033, USA.
Introduction:
Radiocapitellar joint arthroplasty is a commonly performed procedure, which often leads to early failure or instability. Few studies assess the effect of radiocapitellar joint arthroplasty on the ulnohumeral joint. We hypothesized that static forces of contact (compressing cartilage, or cartilage relaxation contact force) would reveal the effect of varying radial head implant size and elbow position on the ulnohumeral joint.
Methods:
A minimally-invasive method of measuring cartilage relaxation contact force was utilized in 10 fresh-frozen human cadaveric specimens that did not require significant dissection or intraarticular sensor placement. Specimens were rigidly fixed in various positions of elbow flexion and forearm pronosupination with increasing radial head implant lengths. Uniaxial distracting forces were applied and displacement was repeatedly measured with resultant best-fit polynomial curves to determine inflections corresponding to the force required to overcome static cartilage relaxation as in previous work.
Findings:
Baseline mean (intra-cadaver) cartilage relaxation contact force was 11.8 N (standard error of the mean = 0.3) at 90° of elbow flexion and neutral rotation. There was little variation within specimens (Intraclass correlation coefficient > 0.94). Cartilage relaxation contact force increased at the ulnohumeral joint with radial head implant overstuffing (> 4 mm, P < 0.05) and elbow flexion (120°, P < 0.001). Pronosupination altered cartilage relaxation contact force in an implant-length independent manner (P < 0.05).
Interpretation:
Radiocapitellar joint arthroplasty implant length and elbow joint position independently contribute to increased cartilage relaxation contact force at the ulnohumeral joint. This further supports attempts at anatomic reconstruction of the radiocapitellar joint to prevent pathologic ulnohumeral joint loading.
Related Concept Videos
Bones of the Upper Limb: Radius
The radius has a nail-shaped head, and a...
Bones of the Upper Limb: Ulna
Bearings: Problem Solving
Bones of the Upper Limb: Humerus
General Case of Eccentric Axial Loading
Consider a member subjected to equal and opposite forces that are applied along a line that does not coincide with the member's neutral axis. In unsymmetrical...
Normal Strain under Axial Loading