Magnetic Resonance Derived Myocardial Strain Assessment Using Feature Tracking
Published on: February 12, 2011
Comparison between conventional and compressed sensing cine cardiovascular magnetic resonance for feature tracking
Tomoyuki Kido1, Kuniaki Hirai2, Ryo Ogawa3
1Department of Radiology, Ehime University Graduate School of Medicine, Hitsukawa, Toon, Ehime, 791-0295, Japan. tomozo0421@gmail.com.
Background:
Feature tracking (FT) has become an established tool for cardiovascular magnetic resonance (CMR)-based strain analysis. Recently, the compressed sensing (CS) technique has been applied to cine CMR, which has drastically reduced its acquisition time. However, the effects of CS imaging on FT strain analysis need to be carefully studied. This study aimed to investigate the use of CS cine CMR for FT strain analysis compared to conventional cine CMR.
Methods:
Sixty-five patients with different left ventricular (LV) pathologies underwent both retrospective conventional cine CMR and prospective CS cine CMR using a prototype sequence with the comparable temporal and spatial resolution at 3 T. Eight short-axis cine images covering the entire LV were obtained and used for LV volume assessment and FT strain analysis. Prospective CS cine CMR data over 1.5 heartbeats were acquired to capture the complete end-diastolic data between the first and second heartbeats. LV volume assessment and FT strain analysis were performed using a dedicated software (ci42; Circle Cardiovasacular Imaging, Calgary, Canada), and the global circumferential strain (GCS) and GCS rate were calculated from both cine CMR sequences.
Results:
There were no significant differences in the GCS (- 17.1% [- 11.7, - 19.5] vs. - 16.1% [- 11.9, - 19.3; p = 0.508) and GCS rate (- 0.8 [- 0.6, - 1.0] vs. - 0.8 [- 0.7, - 1.0]; p = 0.587) obtained using conventional and CS cine CMR. The GCS obtained using both methods showed excellent agreement (y = 0.99x - 0.24; r = 0.95; p < 0.001). The Bland-Altman analysis revealed that the mean difference in the GCS between the conventional and CS cine CMR was 0.1% with limits of agreement between -2.8% and 3.0%. No significant differences were found in all LV volume assessment between both types of cine CMR.
Conclusion:
CS cine CMR could be used for GCS assessment by CMR-FT as well as conventional cine CMR. This finding further enhances the clinical utility of high-speed CS cine CMR imaging.
More Related Videos
11:13Quantification of Mouse Heart Left Ventricular Function, Myocardial Strain, and Hemodynamic Forces by Cardiovascular Magnetic Resonance Imaging
Published on: May 24, 2021
08:10Estimating Bilateral Atrial Function by Cardiovascular Magnetic Resonance Feature Tracking in Patients with Paroxysmal Atrial Fibrillation
Published on: July 20, 2022
Related Concept Videos
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...