Digital Hemoglobinometers as Point-of-Care Testing Devices for Hemoglobin Estimation: A Validation Study from India
Kapil Yadav1, Shashi Kant1, Gomathi Ramaswamy1
1National Centre of Excellence and Advanced Research on Anemia Control, Centre for Community Medicine, All India Institute of Medical Sciences, New Delhi, India.
Introduction:
India has launched Anemia Mukt Bharat (AMB) strategy in 2018, to deal with the high burden of anemia in the country. Point-of-care testing (POCT) of anemia using digital hemoglobinometers and treatment is one of the primary interventions under AMB. This study aimed to determine the diagnostic validity of digital hemoglobinometers (TrueHb and HemoCue 301) for screening of anemia compared to hematology analyzer.
Methodology:
A hospital-based, cross-sectional study was conducted among pregnant women attending antenatal clinics of a selected primary health center and subdivisional hospital of Haryana, India, during January 2019. Hemoglobin (Hb) levels of the pregnant women were estimated in digital hemoglobinometers using capillary blood samples and hematology analyzer using venous blood samples. Bias, limits of agreement (LOA), and validity of digital hemoglobinometers were assessed against a hematology analyzer.
Results:
A total of 110 pregnant women were included. Bias (LOA) in Hb values estimated using digital hemoglobinometers was -0.09 g/dL (-1.97 to 1.80 ) for HemoCue301 and -0.04 g/dL (-1.69 to 1.60) for TrueHb compared to the hematology analyzer. HemoCue® 301 (sensitivity: 86% and specificity: 83%) had relatively higher sensitivity and specificity compared to TrueHb (sensitivity: 78.9% and specificity: 81%).
Conclusions:
Digital hemoglobinometers have high sensitivity and specificity. Thus, these can be a potential POCTs for screening of anemia in peripheral health facilities. Further studies are required to establish the validity of the digital hemoglobinometers at community settings.
More Related Videos
09:33Visualizing Field Data Collection Procedures of Exposure and Biomarker Assessments for the Household Air Pollution Intervention Network Trial in India
Published on: December 23, 2022
07:24A Precision Medicine Tool for Measurement and Monitoring of Hemoglobin S in Sickle Cell Disease Patients Receiving Transfusion Therapy
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Assessment of blood pressure in brachial artery(two-step method)
Blood Studies I: ABG and VBG
Arterial Blood Gas (ABG)
Arterial Blood Gas (ABG) studies are crucial for assessing the lungs' ability to supply oxygen and remove carbon dioxide, reflecting the patient's ventilation status. They also help understand the kidneys' capacity to...