Radiation dose evaluation of pediatric patients in CT brain examination: multi-center study

Fotso Kamdem Eddy1, Samba Odette Ngano2,3, Fotue Alain Jervé2

  • 1Unité de Recherche de la Matière Condensée, d'Electronique et de Traitement du Signal, Department of Physics, Faculty of Science, University of Dschang, Dschang, Cameroon. eddyfotsokamdem@yahoo.fr.

Scientific Reports
|February 26, 2021
PubMed

Insights

Cameroon lacks pediatric radiation dose standards. A study found CT scans in Cameroonian hospitals significantly exceed French Diagnostic Reference Levels (DRL), necessitating protocol optimization for pediatric radiation protection.

Area of Science:

  • Medical Physics
  • Radiology
  • Radiation Protection

Background:

  • Cameroon currently lacks established Pediatric Regulatory Diagnostic Reference Levels (DRL).
  • Diagnostic Reference Levels (DRLs) are crucial for optimizing radiation exposure and ensuring patient safety in medical imaging.
  • France utilizes DRLs to standardize and optimize examination protocols.

Purpose of the Study:

  • To evaluate radiation dose and acquisition parameters in pediatric cranial CT scans in Cameroon.
  • To compare these parameters against France's DRL to identify potential optimization needs.
  • To propose the development of DRLs for pediatric CT in Cameroon.

Main Methods:

  • A retrospective, multicenter study involving 320 pediatric patients undergoing cranial CT scans.
  • Analysis of CT acquisition parameters: tube potential (kV), tube current (mA), slice thickness (T), scanning techniques, CTDIvol, and DLP.
  • Comparison of the 75th percentile dose values with France's DRL.

Main Results:

  • Pediatric CTDIvol values in Cameroonian centers were higher than France's DRL across all age groups (1-4, 5-9, 10-14 years).
  • Dose Length Product (DLP) values also showed a similar trend of being higher than French DRLs.
  • A significant dosimetric overshoot was observed, indicating a need for protocol adjustments.

Conclusions:

  • Cameroonian pediatric CT protocols result in higher radiation doses compared to established international benchmarks (France's DRL).
  • Urgent need to implement DRLs and optimize protocols to ensure radiation protection for pediatric patients.
  • Establishment of a quality control structure is recommended for consistent application of radiation safety principles in Cameroon.