Related Experiment Video
Updated: Nov 16, 2025

Development of a Lateral Flow Immunochromatographic Strip for Rapid and Quantitative Detection of Small Molecule Compounds
Published on: November 13, 2021
Lateral flow immunogold assay as a rapid detection tool for screening of congenital hypothyroidism
Younes Sbihi1, Najia El Abbadi1, Abdelghani Iddar1
1Biotechnology and Engineering of Biomolecules Unit, National Centre for Nuclear Energy, Science and Technology (Cnesten-morocco), Rabat, Morocco.
Insights
A new rapid test accurately detects thyroid-stimulating hormone (TSH) for congenital hypothyroidism (CH) screening. This lateral flow assay offers a visual detection limit of 10 μUI/mL, aiding early diagnosis in underserved areas.
Area of Science:
- Biochemistry
- Immunology
- Medical Diagnostics
Background:
- Congenital hypothyroidism (CH) is a leading preventable cause of mental retardation.
- Newborn screening programs commonly use thyroid-stimulating hormone (TSH) tests for CH diagnosis.
Purpose of the Study:
- To develop a rapid, sensitive immunochromatographic lateral flow assay for TSH detection.
- To evaluate the assay's performance for congenital hypothyroidism screening.
Main Methods:
- A sandwich lateral flow immunoassay utilizing monoclonal antibodies (MAbs) conjugated to colloidal gold nanoparticles.
- Optimization of colloidal gold-MAbs conjugation and determination of TSH sensitivity (5-50 μUI/mL).
- Comparison of assay results with radioimmunoassay (RIA) using umbilical cord serum samples.
Main Results:
- The developed assay demonstrated a visual limit of detection of 10 μUI/mL for TSH.
- Color intensity correlated directly with TSH concentration.
- Results from the lateral flow assay showed agreement with conventional RIA methods.
Conclusions:
- The rapid lateral flow immunoassay is a viable tool for initial TSH screening.
- This assay has potential for effective congenital hypothyroidism screening, particularly in resource-limited or rural settings.
Abstract:
Congenital hypothyroidism (CH) is one of the most common preventable causes of mental retardation. The majority of infants are diagnosed after detection through newborn screening programs using thyroid-stimulating hormone (TSH) test. A rapid immunochromatographic lateral flow assay based on monoclonal antibodies (MAbs) colloidal gold nanoparticles was developed in a sandwich format for the detection of TSH. Two MAbs binding distinct TSH epitopes are used; one is conjugated to the detection reagent while the other is immobilized at the test line on the membrane. The colloidal gold was prepared by the reduction of gold salt coupled with MAbs and this optimal concentration was determined by spectrophotometry method. The sensitivity of our developed lateral flow immunoassay was determined using 5, 10, 15, 25 and 50 μUI/mL of TSH. The color intensity of the test line was directly proportional to the TSH concentration and the visual limit of detection was 10 μUI/mL. Twenty samples of umbilical cord serum were analyzed by the developed strips and the intensity of the signal was in agreement with the results obtained by the conventional radioimmunoassay method. The results suggest that this rapid test can be used in initial screening for congenital hypothyroidism especially in rural areas.
More Related Videos
08:01Detection of Invasive Pulmonary Aspergillosis in Haematological Malignancy Patients by using Lateral-flow Technology
Published on: March 22, 2012
06:50A High-Throughput Electrochemiluminescence 7-Plex Assay Simultaneously Screening for Type 1 Diabetes and Multiple Autoimmune Diseases
Published on: May 29, 2020