Related Experiment Videos
[Temperature distribution in interstitial hyperthermia]
H von Lieven1, G Schwingen, R Proske
1Wilhelm-Conrad-Röntgen-Klinik, Universität Giessen.
Abstract:
The temperature distribution during the application of interstitial hyperthermia was measured in a static and a dynamic phantom. Despite the temperature-controlled release of energy over six electrodes, there were considerable temperature heterogeneities in the target volume. The temperatures measured in partial volumes were largely dependent on the regional perfusion.