Quantitative Susceptibility Mapping Using a Multispectral Autoregressive Moving Average Model to Assess Hepatic Iron
Aaryani Tipirneni-Sajja1,2, Ralf B Loeffler1,3, Jane S Hankins4
1Department of Diagnostic Imaging, St. Jude Children's Research Hospital, Memphis, Tennessee, USA.
Journal of Magnetic Resonance Imaging : JMRI
|February 26, 2021
Summary
This study introduces ARMA-QSM, a novel method for accurately assessing hepatic iron content (HIC) by overcoming limitations of R2*-MRI. It provides a confounder-free approach for evaluating iron overload.
Area of Science:
- Medical Imaging
- Quantitative MRI
- Biomarkers
Background:
- R2*-MRI is used for hepatic iron content (HIC) assessment but is affected by coexisting pathologies like fibrosis and steatosis.
- These pathologies can confound HIC estimations, potentially impacting iron chelation therapy guidance.
Purpose of the Study:
- To evaluate the efficacy of a multispectral autoregressive moving average (ARMA) model combined with quantitative susceptibility mapping (QSM).
- To determine if ARMA-QSM can provide a confounder-free measurement of magnetic susceptibility for HIC prediction.
Main Methods:
- Utilized a phantom study and an in vivo cohort of 48 patients.
- Performed ARMA-QSM modeling on 3D multi-echo gradient echo (GRE) signals to estimate R2*, fat fraction (FF), and susceptibility.
- Calculated R2*-based dry HIC using 2D GRE acquisition and a published calibration curve.
Main Results:
- ARMA R2* and susceptibility values strongly correlated with iron concentrations in phantoms (R² ≥ 0.9).
- In patients, ARMA R2* highly correlated with clinical HIC (R² = 0.97), and susceptibility showed good correlation (R² = 0.82).
- ARMA-QSM simultaneously estimated susceptibility-based wet HIC, R2*-based dry HIC, and FFs from a single acquisition.
Conclusions:
- ARMA-QSM demonstrates feasibility for simultaneously estimating multiple parameters from a single multi-echo GRE scan.
- Both R2*- and susceptibility-based HIC values from ARMA-QSM correlate well with clinical dry HIC.
- ARMA-QSM offers a promising approach for confounder-free assessment of hepatic iron overload.