The Becton Dickinson Barricor Tube Does Not Introduce Isobaric Interferences in Mass Spectrometry-based Measurements
1Department of Pathology and Laboratory Medicine, Weill Cornell Medical College, New York, NY.
Background:
Testosterone analysis by mass spectrometry is essential for accurate low-level results. Unfortunately, the commonly used gel-based separator tubes introduce an intense, isobaric interference that can coelute with testosterone and impede analysis. The new Becton Dickinson (BD) BarricorTM tube (BT) uses an elastomer separator that might not introduce such interferences.
Methods:
Testosterone concentrations and chromatograms were obtained for plasma samples from plasma separator tubes (PSTs) and the new BTs using a laboratory-developed liquid chromatography-tandem mass spectrometry (LC-MS/MS) method.
Results:
Intense signals close to the testosterone retention time were observed in the testosterone quantifier transition (289.2→97.0) from the PSTs; resolving these signals required longer chromatographic run times. No isobaric interferences were introduced by the BTs, allowing for shorter chromatographic run times and faster inject-to-inject times.
Conclusion:
BTs do not introduce the interferences observed in PSTs in LC-MS/MS testosterone analysis. These tubes can be used as a single tube that both separates cells and is amenable to LC-MS/MS testosterone analysis without additional chromatographic run times.
Related Concept Videos
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Tandem Mass Spectrometry
Secondary fragmentations occur in the interaction cell and can be induced by various factors. Fragmentation induced by collision with inert gases, such as N2, Ar, He, etc., is called collision-induced...
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Mass Spectrometry: Isotope Effect
High-Resolution Mass Spectrometry (HRMS)
Mass Analyzers: Common Types