Non-invasive monitoring of carboxyhemoglobin during hyperbaric oxygen therapy
Daniel Bidstrup1, Frederikke Ravn1, Isabel Guida Smidt-Nielsen1
1Department of Anaesthesia, Hyperbaric Unit, Centre of Head and Orthopaedics, Copenhagen University Hospital, Rigshospitalet, Copenhagen, Denmark.
Introduction:
This study aimed to assess the capability of a pulse CO-oximeter to continuously monitor carboxyhemoglobin (COHb) during hyperbaric oxygen (HBO2) therapy. We estimated limits of agreement (LOA) between blood gas analysis and pulse CO-oximeter for COHb during HBO2 therapy in patients suffering from acute CO poisoning. Furthermore, we did a medicotechnical evaluation of the pulse CO-oximeter in hyperbaric conditions.
Methods:
We conducted a prospective, non-clinical, observational study in which we included n=10 patients with acute CO poisoning referred for HBO2 therapy. We did five repeated measurements of COHb for each patient during the HBO2 therapy. Bland-Altman analysis for multiple observations per individual was used to assess the agreement. The a priori LOA was ±6% for COHb. For the medicotechnical evaluation continuous measurements were obtained throughout each complete HBO2 therapy. The measurements were visually inspected and evaluated.
Results:
The Bland-Altman analysis showed that the pulse CO-oximeter overestimated COHb by 2.9 % [±1.0%] and the LOA was ±7.3% [±1.8%]. The continuous measurements by pulse CO-oximetry showed fluctuating levels of COHb and summarized saturations reached levels above 100%. Measurements were not affected by changes in pressure.
Conclusion:
To our knowledge, this study is the first to assess LOA and demonstrate use of a non-invasive method to measure COHb during HBO2 therapy. The pulse CO-oximeter performed within the manufactures reported LOA (±6%) despite hyperbaric conditions and was unaffected by changes in pressure. However, summarized saturations reached levels above 100%.
More Related Videos
Related Concept Videos
Oxygen Transport in the Blood
Hypoxia
Types of Hypoxia
There are four primary types of hypoxia, each resulting from a different cause:
1. Anemic hypoxia: This type occurs due to insufficient oxygen delivery caused by a lack of red blood cells (RBCs) or RBCs with abnormal or...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Blood Studies I: ABG and VBG
Arterial Blood Gas (ABG)
Arterial Blood Gas (ABG) studies are crucial for assessing the lungs' ability to supply oxygen and remove carbon dioxide, reflecting the patient's ventilation status. They also help understand the kidneys' capacity to...