Different impact factors for accurate oscillometric blood pressure measurement between sinus rhythm and atrial
Feng Xie1,2, Jinsong Xu1, Huan Liu1
1Department of Cardiovascular Medicine, The Second Affiliated Hospital of Nanchang University, Nanchang, China.
Insights
Oscillometric blood pressure (BP) devices are less accurate in patients with atrial fibrillation (AF) compared to those in sinus rhythm (SR). Ventricular rate significantly impacts BP measurement accuracy in AF patients.
Area of Science:
- Cardiology
- Biomedical Engineering
- Medical Devices
Background:
- Accurate blood pressure (BP) monitoring is crucial in cardiovascular care.
- Oscillometric BP measurement is widely used but its accuracy can be affected by arrhythmias.
- Atrial fibrillation (AF) and sinus rhythm (SR) represent different cardiac electrical activities that may influence BP readings.
Purpose of the Study:
- To compare the accuracy of upper arm oscillometric BP measurements against intra-aortic BP measurements in patients with AF and SR.
- To investigate the influence of patient-specific factors like age and ventricular rate (VR) on oscillometric BP accuracy in both groups.
Main Methods:
- Simultaneous measurements of upper arm oscillometric BP and intra-aortic BP were taken three times in 124 AF patients and 98 SR patients undergoing coronary arteriography.
- The difference (delta BP) and percentage difference (PD-BP) between the two measurement methods were calculated.
- Correlation analyses were performed to assess the impact of age and VR on BP measurement accuracy.
Main Results:
- Oscillometric BP underestimated intra-aortic BP by 2.2/1.6 mmHg (SBP/DBP) in SR patients, but significantly more by 5.8/4.8 mmHg in AF patients.
- In SR patients, age showed correlations with delta SBP and delta DBP.
- In AF patients, ventricular rate (VR) was negatively correlated with delta SBP and delta DBP, indicating decreased accuracy with increased VR.
Conclusions:
- Oscillometric BP measurement accuracy is significantly lower in patients with AF compared to SR.
- Ventricular rate is a key factor affecting oscillometric BP accuracy in AF patients.
- Age may influence oscillometric BP accuracy in SR patients, while VR is a critical determinant in AF patients.
Abstract:
This study was conducted to evaluate whether the impact factor on the accuracy of oscillometric blood pressure (BP) is different between the patients with atrial affiliation (AF) and sinus rhythm (SR). This study included 124 AF patients and 98 SR patients who underwent coronary arteriography. Upper arm oscillometric and intro-aortic BP were simultaneously measured three times, and the averages were recorded as final values, respectively. The difference between oscillometric and intro-aortic BP was recorded as delta BP. For reducing the influence of baseline BP on delta BP, a percentage difference (PD-BP) was calculated with the formula: PD-BP = delta BP/intro-aortic BP × 100%. The oscillometric SBP and DBP levels slightly underestimated intro-aortic BP (by 2.2/1.6 mmHg) in the SR group, but significantly underestimated (by 5.8/4.8 mmHg) in the AF group. Age was negatively correlated with delta SBP and positively with delta DBP in SR patients, but not correlated in AF patients. Otherwise, ventricular rate (VR) was negatively correlated with delta SBP or delta DBP in AF, which indicated that the accuracy of oscillometric BP declined as the VR increased. With intro-aortic BP as reference, the accuracy of oscillometric BP in AF patients is lower than SR patients. Meanwhile, age may impact the oscillometric BP accuracy in SR patients, but VR may impact that in AF patients.
More Related Videos
Related Concept Videos
Measurement of Blood Pressure
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Errors occurring during blood pressure monitoring
Several factors...
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)