Related Experiment Video
Updated: Nov 15, 2025

Synthesis of Soft Polysiloxane-urea Elastomers for Intraocular Lens Application
Published on: March 8, 2019
Factory-Made Elastic Silicone Sheet as an Alternative for the Posterior Capsule of the Lens During
Rinko Akamine1, Naomi Miyamoto1, Shoko Kiritoshi1
1Osaka Red Cross Hospital, Osaka, Japan.
Background:
We previously reported a novel technique using an elastic silicone bowl, which was handmade from a phaco test chamber, to salvage dislocated lenses and perform phacoemulsification and aspiration (PEA) on the bowl. However, creating the bowl was challenging.
Methods:
A factory-made elastic silicone sheet (ESS), a ready-to-use alternative for the posterior capsule of the lens, was prepared. We used this sheet to salvage and support the dislocated/subluxated lenses in 11 eyes of nine patients.
Results:
We could use a sheet immediately after we opened the package. In all the cases, we could insert a sheet beneath the lens and PEA performed at the center of the operative field. There was only one incident where the lens could not be placed onto the disc at the first attempt, but the lens was eventually salvaged. There was no risk of rupture during PEA, and the ESS could be removed completely without any complications.
Conclusion:
The factory-made ESS described here can be used for salvaging dislocated/subluxated lenses and safely performing PEA.