Computed tomography with a full FOV photon-counting detector in a clinical setting, the first experience
Jiří Ferda1, Tomáš Vendiš1, Thomas Flohr2
1Department of the Imaging, University Hospital Pilsen, Alej Svobody 80, 323 00, Pilsen, Czechia.
Objectives:
to assess the feasibility of CT with an integrated photon-counting-detector system (PC-CT) in the body imaging of clinical patients.
Methods:
120 examinations using photon counting detector CT were evaluated in six groups: 1/ a standard-dose lung, 2/ low-dose lung, 3/ ultra-high resolution (UHR) lung, 4/ standard-dose abdominal, 5/ lower-dose abdominal, 6/ UHR abdominal CTA. All CT examinations were performed on a single-source prototype device equipped with a photon counting detector covering a 50 cm scan field of view. Standard dose examinations were performed with the use of detector element size of 0.4 mm, ultra-high-resolution examinations with the detector element size of 0.2 mm, respectively. The stability of the system during imaging was tested. The diagnostic quality of the acquired images was assessed based on the imaging of key structures and the noise level in five-point scale, the effective dose equivalent, dose length product and noise level, and also relation to body mass index and body surface area were compared with three similar groups of CT images made with energy integrating high end scanner. The parameters were evaluated using Wilcoxon test for independent samples, the independence was tested using Kruskal-Wallis test.
Results:
When PC-CT images radiation dose is compared with the similar imaging using energy integrating CT, the PC-CT shows lower dose in ultra-high resolution mode, the dose is significantly lower (p < 0.0001), the standard dose examinations were performed with the comparable radiation doses. PC-CT exhibited the significantly higher ratio between parenchyma signal and background noise both in lung and in abdominal imaging (p < 0.0001).
Conclusions:
PC-CT showed imaging stability and excellent diagnostic quality at dose values that are comparable or better to the quality of energy integrating CT, the better signal and improved resolution is most important advantage of photon counting detector CT over energy integrating detector CT.
More Related Videos
08:36Multi-Tracer Studies of Brain Oxygen and Glucose Metabolism Using a Time-of-Flight Positron Emission Tomography-Computed Tomography Scanner
Published on: June 7, 2024
11:49Using High Resolution Computed Tomography to Visualize the Three Dimensional Structure and Function of Plant Vasculature
Published on: April 5, 2013
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...