Prostatic Artery Embolization: Influence of Cone-Beam Computed Tomography on Radiation Exposure, Procedure Time, and
F Bürckenmeyer1, I Diamantis1, T Kriechenbauer1
1Department for Diagnostic and Interventional Radiology, University Hospital Jena, Am Klinikum 1, 07747, Jena, Germany.
Purpose:
To evaluate the effect of cone-beam computed tomography (CBCT) on radiation exposure, procedure time, and contrast media (CM) use in prostatic artery embolization (PAE).
Materials And Methods:
Seventy-eight patients were enrolled in this retrospective, single-center study. All patients received PAE without (group A; n = 39) or with (group B; n = 39) CBCT. Total dose-area product (DAPtotal; Gycm2), total entrance skin dose (ESDtotal; mGy), and total effective dose (EDtotal; mSv) were primary outcomes. Number of digital subtraction angiography (DSA) series, CM use, fluoroscopy time, and procedure time were secondary outcomes. PAE in group A was performed by a single radiologist with 15 years experience, PAE in group B was conducted by four radiologists with 4 to 6 years experience.
Results:
For groups A vs. B, respectively, median (IQR): DAPtotal 236.94 (186.7) vs. 281.20 (214.47) Gycm2(p = 0.345); EDtotal 25.82 (20.35) vs. 39.84 (23.75) mSv (p = < 0.001); ESDtotal 2833 (2278) vs. 2563 (3040) mGy(p = 0.818); number of DSA series 25 (15) vs. 23 (10)(p = 0.164); CM use 65 (30) vs. 114 (40) mL(p = < 0.001); fluoroscopy time 23 (20) vs. 28 (25) min(p = 0.265), and procedure time 70 (40) vs.120 (40) min(p = < 0.001). Bilateral PAE was achieved in 33/39 (84.6%) group A and 32/39 (82.05%) group B(p = 0.761), all other patients received unilateral PAE. There were no significant differences between clinical parameters and origins of the prostatic arteries (PA) (p = 0.206-1.00).
Conclusion:
Operators with extensive expertise on PAE may not benefit from addition of CBCT to DSA runs, whereas for operators with less expertise, CBCT when used alongside with DSA runs increased the overall radiation exposure.
More Related Videos
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
Related Concept Videos
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation I: X-ray and CT
Imaging Studies VII: Vascular Imaging
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...