Helium-assisted enhanced discharge in a hollow needle for high-field asymmetric ion mobility spectrometry (FAIMS)
Hua Li1, Xiaoxia Du2, Hongda Zeng2
1School of Life and Environmental Sciences, Guilin University of Electronic Technology, Guilin, 541004, Guangxi, China. lihua@guet.edu.cn.
Abstract:
A carrier gas mixture of nitrogen and helium has been employed to improve the resolving power at the expense of sensitivity for planar high-field asymmetric ion mobility spectrometry (FAIMS) in previous work. In this paper, a new hollow needle-to-ring ion source was developed, where the helium and nitrogen enter from the hollow needle and ring, respectively. It was found that the signal strengths of acetone, ethanol, and ethyl acetate increased by 8.5, 2.0, and 3.3 times for helium ratios of 20%, 20%, and 10%, respectively. At the same time, the absolute value of compensation voltage and the number of ion peaks increases. It shows that adding an appropriate helium ratio to nitrogen simultaneously improved the sensitivity and resolving power of planar FAIMS, which is reported for the first time.
Related Concept Videos
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Atomic Emission Spectroscopy: Lab
Chemical Ionization (CI) Mass Spectrometry
Atomic Emission Spectroscopy: Overview