Potentiometric Response of Solid-State Sensors Based on Ferric Phosphate for Iron(III) Determination
Andrea Paut1, Ante Prkić1, Ivana Mitar2
1Department of Analytical Chemistry, Faculty of Chemistry and Technology, University of Split, R. Boškovića 35, 21000 Split, Croatia.
Abstract:
A novel ion-selective electrode with membranes based on iron(III) phosphate and silver sulfide integrated into a completely new electrode body design has been developed for the determination of iron(III) cations. The best response characteristics with linear potential change were found in the iron(III) concentration range from 3.97 × 10-5 to 10-2 mol L-1. The detection limit was found to be 2.41 × 10-5 mol L-1 with a slope of -20.53 ± 0.63 and regression coefficient of 0.9925, while the quantification limit was 3.97 × 10-5 M. The potential change per concentration decade ranged from -13.59 ± 0.54 to -20.53 ± 1.56 for Electrode Body 1 (EB1) and from -17.28 ± 1.04 to -24 ± 1.87 for Electrode Body 2 (EB2), which is presented for the first time in this work. The prepared electrode has a long lifetime and the ability to detect changes in the concentration of iron cations within 20 s. Membrane M1 showed high recoveries in the determination of iron cations in iron(III) standard solutions (98.2-101.2%) as well as in two different pharmaceuticals (98.6-106.5%). This proves that this type of sensor is applicable in the determination of ferric cations in unknown samples, and the fact that all sensor parts are completely manufactured in our laboratory proves the simplicity of the method.
Related Concept Videos
Potentiometry: Membrane Electrodes
Potentiometry: Types of Electrodes
The Standard Hydrogen Electrode (SHE) is a widely used reference electrode that maintains zero potential across all temperatures. However, its need for a continuous hydrogen gas supply renders it impractical for everyday use.
An alternative to SHE is the Saturated Calomel Electrode (SCE). This electrode features an...
Potentiometric Titration: Overview
Precipitation Titration: Endpoint Detection Methods
In the Volhard method, a standard excess of AgNO3 is first added to the...
Potentiometry: Overview
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...