Titanium and Platinum-Fluoroplastic Stapes Prostheses Visualization on Cone Beam Computed Tomography and
Valerie Dahm1, Ursula Schwarz-Nemec2, Alice B Auinger1
1Department of Otorhinolaryngology, Head and Neck Surgery, Medical University of Vienna, 1090 Vienna, Austria.
Life (Basel, Switzerland)
|March 6, 2021
Summary
Cone beam computed tomography (CBCT) offers a potential advantage over computed tomography (CT) for visualizing stapes prostheses after surgery. This imaging technique may reduce metal artifacts, aiding in the evaluation of hearing loss or vertigo in patients.
Area of Science:
- Otolaryngology
- Medical Imaging
- Biomaterials
Background:
- Stapes prostheses are crucial for hearing restoration after stapedotomy.
- Imaging challenges arise due to metal artifacts from prostheses, impacting diagnosis for hearing loss or vertigo.
- Cone beam computed tomography (CBCT) is increasingly recognized for its potential in reducing metal artifacts.
Purpose of the Study:
- To compare the visualization accuracy of stapes prostheses using CBCT versus traditional CT.
- To evaluate the impact of prosthesis material (platinum-fluoroplastic vs. titanium) on imaging quality.
- To assess the utility of CBCT in identifying prosthesis position and potential complications like vestibular intrusion.
Main Methods:
- Retrospective analysis of 41 patients (45 implanted ears) with platinum-fluoroplastic or titanium stapes prostheses.
- Comparison of imaging data from CT and CBCT, focusing on prosthesis visualization, length, and position.
- Correlation of imaging findings with intraoperative data and audiologic results.
Main Results:
- Functional length measurements showed a mean error of -0.17 mm (±0.20).
- Platinum-fluoroplastic prostheses were significantly underestimated in length compared to titanium.
- A blinded neuroradiologist misidentified the position of two platinum-fluoroplastic prostheses due to underestimation of length and poor radiodensity.
Conclusions:
- Surgeons and radiologists must recognize the varying radiologic characteristics of different stapes prosthesis materials.
- CBCT presents a viable imaging alternative for post-stapes surgery evaluation, potentially offering reduced radiation exposure.
- CBCT may improve the assessment of prosthesis position in patients experiencing vertigo or hearing loss after stapedotomy.
More Related Videos
Related Concept Videos
Computed Tomography
7.6K
Tomography refers to imaging by sections. Computed tomography (CT) is a non-invasive imaging technique that uses computers to analyze several cross-sectional X-rays to reveal minute details about structures in the body.
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
7.6K
Imaging Studies III: Computed Tomography
126
DefinitionComputed Tomography (CT) of the genitourinary (GU) tract is a non-invasive imaging modality that utilizes X-rays and computer processing to generate detailed cross-sectional images of the urinary system, encompassing the kidneys, ureters, bladder, and adjacent structures such as the adrenal glands.PurposeCT scans of the GU tract serve several diagnostic and therapeutic purposes, including:Diagnosis of Urinary Tract Diseases: Detects kidney stones, tumors, cysts, and congenital...
126