Background:
Aseptic tibial loosening is a frequent cause of long-term failure following primary cemented total knee replacement. Failure of the tibial implant can occur at the implant-cement interface or at the cement-bone interface. Currently, it is unknown at which interface failure occurs in cases of aseptic tibial loosening. The following study was designed to determine which interface represents the "weak link" for tibial implant fixation.
Methods:
We performed a retrospective analysis of 149 patients who were revised secondary to aseptic tibial loosening at our institution from 2005 to 2017. Operative reports and radiographs were reviewed on each patient to determine the location and pattern of fixation failure.
Results:
Implant failure was more prevalent at the implant-cement than cement-bone interface, 140/149 (94.0%) vs 9/149 (6.0%). Additionally, we noted 2 distinct patterns of failure in patients that loosened at the implant-cement interface. Ninety of 140 (64.3%) patients developed varus collapse pattern of failure. Forty-nine of 140 (35.0%) patients developed failure between the implant-cement interface without angulation. All 149 patients had heterotopic bone formation anterior to the tibial baseplate, which was consistent regardless of which interface failed.
Conclusion:
The most frequent interface failure identified in our study was at the implant-cement interface, 140/149 (94.0%). This finding has substantial clinical ramifications. Because failure was predominantly at the implant-cement interface there may be design opportunities for increasing implant fixation to cement. Implants with improved undersurface tibial tray features may be necessary to mitigate the risk of failure at this interface, especially in overly active patients or those with elevated body mass indices.
More Related Videos
08:20A Reliable and Reproducible Critical-Sized Segmental Femoral Defect Model in Rats Stabilized with a Custom External Fixator
Published on: March 24, 2019
04:41Treatment with Locking Intramedullary Nailing for Intertrochanteric Fracture of the Femur Utilizing a New Awl with a Distal Positioner
Published on: June 6, 2025
Related Concept Videos
Knee Joint
A total of seven ligaments support the knee joint. The patellar ligament, which is also attached to the quadriceps femoris...
Bones of the Lower Limb: Femur and Patella