Related Experiment Video
Updated: Nov 14, 2025

An Intestine/Liver Microphysiological System for Drug Pharmacokinetic and Toxicological Assessment
Published on: December 3, 2020
Determination and validation of aprepitant in rat plasma using LC-MS/MS
Nazlı Erdoğar1, Tuba Reçber2, Alper Bektas İskit3
1Department of Pharmaceutical Technology, Hacettepe University, Faculty of Pharmacy, Sıhhiye-Ankara 06100, Turkey.
Abstract:
Aim: The assessment of efficacy should be paralleled with extensive pharmacokinetic parameters, and a valid bioanalytical method is a pre-condition for accurate plasma concentration. Materials & methods: A simple, specific, rapid and sensitive LC-MS/MS method has been developed for quantitative analysis of aprepitant in rat plasma. A C18 column was used as stationary phase and the mobile phase consisted of a mixture of formic acid in water and formic acid in acetonitrile. Quantification was performed using multiple reaction monitoring mode. Results: The selectivity, linearity, accuracy, precision, robustness and ruggedness of the method were evaluated in accordance with bioanalytical method validation guideline of ICH and all results were within the acceptable range. Conclusion: The validated LC-MS/MS method was found to be useful for the quantitative analysis of aprepitant in rat plasma samples.