Dose quantities for measurement and comparison of doses to individual patients in computed tomography (CT)
Colin J Martin1, Abdullah Abuhaimed2, Choonsik Lee3
1Department of Clinical Physics and Bioengineering, University of Glasgow, Gartnavel Royal Hospital, Glasgow G12 0XH, United Kingdom.
Abstract:
The dose quantities displayed routinely on CT scanners, the volume averaged CT dose index (CTDIvol) and dose length product, provide measures of doses calculated for standard phantoms. The American Association of Medical Physics has published conversion factors for the adjustment of CTDIvolto take account of variations in patient size, the results being termed size-specific dose estimate (SSDE). However, CTDIvoland SSDE, while useful in comparing and optimising doses from a set procedure, do not provide risk-related information that takes account of the organs and tissues irradiated and associated cancer risks. A derivative of effective dose that takes account of differences in body and organ sizes and masses, referred to here as size-specific effective dose (SED), can provide such information. Data on organ doses from NCICT software that is based on Monte Carlo simulations of CT scans for 193 adult phantoms have been used to compute values of SED for CT examinations of the trunk and results compared with corresponding values of SSDE. Relationships within ±8% were observed between SED and SSDE for scans extending over similar regions for phantoms with a wide range of sizes. Coefficients have been derived from fits of the data to estimate SED values from SSDEs for different regions of the body for scans of standard lengths based on patient height. A method developed to take account of differences in scan length gave SED results within ±5% of values calculated using the NCI phantom library. This approach could potentially be used to estimate SED from SSDE values, allowing their display at the time a CT scan is performed.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Dose Size and Dosing Frequency: Determination Methods
Radiological Investigation I: X-ray and CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...