Aim:
To review the removal of Y-shaped airway self-expanding covered metallic stents using the interventional technique under fluoroscopy.
Materials And Methods:
The clinical data of 33 patients who underwent removal of Y-shaped airway self-expanding covered metallic stents from March 2011 to August 2019 were analysed retrospectively.
Results:
A total of 35 Y-shaped stents were removed. The average indwelling duration of the tracheal stents was 101.7 ± 105.4 days. Four cases were removed via the conventional method (grasping the upper tip of the stent to collapse and adduct the proximal end of the stent), whereas 31 cases were removed using the reversal method (grasping the distal end of the stent to invert and strip out the stent). The duration of stent removal was 24.3 ± 12.4 minutes (median time, 20 minutes).
Conclusion:
The interventional radiology technique is a feasible, safe, and effective method for removing Y-shaped airway self-expanding covered metallic stents, and can be considered for use in the clinical setting.
Related Concept Videos
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Tracheostomy Suctioning II: Procedure
Tracheostomy Decannulation
Description of the Procedure
Decannulation refers to the permanent removal of the tracheostomy tube, signaling the resolution of the condition that initially necessitated the tracheostomy. The process requires a well-coordinated interplay between...