Related Experiment Video
Updated: Nov 14, 2025

The Use of Mixed Reality in Custom-Made Revision Hip Arthroplasty: A First Case Report
Published on: August 4, 2022
The reasons for ceramic-on-ceramic revisions between the third- and fourth-generation bearings in total hip
Sang-Min Kim1, Kee Hyung Rhyu2, Jeong Joon Yoo3
1Department of Orthopedic Surgery, Korea University College of Medicine, Guro Hospital, Seoul, South Korea.
Insights
Revision rates for ceramic-on-ceramic total hip arthroplasty (THA) were significantly lower for fourth-generation compared to third-generation bearings. Bearing-related issues were the primary driver for revisions in CoC THA, particularly in older ceramic designs.
Area of Science:
- Orthopedic Surgery
- Biomaterials Science
- Arthroplasty Research
Background:
- Ceramic-on-ceramic (CoC) bearings are utilized in total hip arthroplasty (THA) to reduce wear debris.
- Understanding revision reasons for CoC THA is crucial for improving implant longevity and patient outcomes.
- Generational differences in CoC bearing technology may influence revision rates and causes.
Purpose of the Study:
- To determine the primary reasons for first revision in CoC total hip arthroplasty (THA).
- To compare revision causes between third-generation and fourth-generation CoC THAs.
- To identify factors associated with bearing-related problems leading to CoC THA revision.
Main Methods:
- Retrospective review of 2194 hips undergoing first revision THA between 2004 and 2013.
- Analysis of 146 hips with CoC bearings requiring revision, categorized by generation (92 third-gen, 54 fourth-gen).
- Statistical analysis to compare revision reasons and identify associated factors.
Main Results:
- The main reasons for CoC THA revision were ceramic fracture and component loosening.
- Bearing-related problems accounted for 28.8% of CoC revisions.
- Third-generation CoC bearings had a significantly higher rate of bearing-related revisions (41.3%) compared to fourth-generation (7.4%) (p < 0.001).
- Younger age, lower ASA grade, and osteonecrosis were linked to bearing-related revisions.
Conclusions:
- Fourth-generation CoC bearings demonstrate a substantially lower incidence of revision due to bearing-related issues compared to third-generation.
- Ceramic fracture and loosening remain significant concerns in CoC THA revisions.
- Patient factors such as younger age, lower ASA grade, and osteonecrosis may predispose to bearing-related complications in CoC THA.
Abstract:
This study aimed to evaluate (1) the overall reasons for first revision in CoC THAs; (2) whether the reasons for revision differ between third-generation and fourth-generation CoC THAs; and (3) the specific factors associated with bearing-related problems as the reason for revision. We retrospectively reviewed 2045 patients (2194 hips) who underwent first revision THA between 2004 and 2013, among which 146 hips with CoC bearings underwent revision. There were 92 hips with third-generation ceramic bearings and 54 hips with fourth-generation ceramic bearings. The major reasons for CoC THA revisions were ceramic fracture and loosening of the cup or stem. When ceramic fracture, squeaking, incorrect ceramic insertion, and unexplained pain were defined as directly related or potentially related to ceramic use, 28.8% (42/146) of CoC revisions were associated with bearing-related problems. Among the third-generation ceramic bearings, revision was performed in 41.3% (38/92) of cases owing to bearing-related problems whereas revisions were performed for only 7.4% (4/54) of cases with fourth-generation ceramic bearings owing to bearing-related problems (p < 0.001). Younger age, lower American Society of Anesthesiologists (ASA) grade, and preoperative diagnosis of osteonecrosis were factors related to CoC THA revisions due to bearing-related problems.
More Related Videos
09:51The Transition to an Anterior-Based Muscle Sparing Approach Improves Early Postoperative Function but is Associated with a Learning Curve
Published on: September 7, 2022
07:33In Vitro Application of a Wireless Sensor in Flexion-Extension Gap Balance of Unicompartmental Knee Arthroplasty
Published on: May 5, 2023