Diagnostic Value of Atrial Fibrillation by Built-in Electrocardiogram Technology in a Blood Pressure Monitor
Keitaro Senoo1,2, Tomonori Miki2, Takashi Okura2
1Department of Cardiac Arrhythmia Research and Innovation, Graduate School of Medical Science, Kyoto Prefectural University of Medicine Kyoto Japan.
Insights
Hypertension and atrial fibrillation (AF) increase stroke risk. A new device combining blood pressure (BP) and electrocardiogram (ECG) monitoring accurately distinguishes AF from sinus rhythm (SR) during BP measurement.
Area of Science:
- Cardiology
- Medical Devices
- Hypertension Research
Background:
- Hypertension in atrial fibrillation (AF) patients is a significant stroke risk factor.
- Accurate differentiation between sinus rhythm (SR) and AF is crucial for patient management.
- Existing monitoring methods may require separate devices for blood pressure (BP) and electrocardiogram (ECG).
Purpose of the Study:
- To evaluate the accuracy of the Complete blood pressure (BP) monitor in differentiating between sinus rhythm (SR) and atrial fibrillation (AF) during BP measurement.
- To assess the device's capability as a single-unit monitor for both BP and ECG readings.
Main Methods:
- Fifty-six patients with persistent AF undergoing catheter ablation were enrolled.
- Simultaneous 12-lead ECGs and Complete device recordings were obtained before and after ablation.
- Complete device interpretations were compared against physician-reviewed ECGs, with cardiologists performing blinded reviews.
Main Results:
- The Complete's automated algorithm achieved 100% sensitivity and 86% specificity in differentiating SR from AF (κ=0.87).
- Physician-interpreted Complete recordings showed 99% sensitivity and 85% specificity (κ=0.85).
- The device demonstrated high accuracy in classifying heart rhythms during BP monitoring.
Conclusions:
- The Complete device, integrating BP and ECG monitoring, accurately differentiates between SR and AF during BP measurement.
- This combined monitoring approach offers a promising tool for managing patients with both hypertension and AF.
- The device's performance supports its potential utility in clinical settings for rhythm assessment.
Abstract:
Hypertension in patients with atrial fibrillation (AF) is a known independent risk factor for stroke. The Complete blood pressure (BP) monitor (Omron Healthcare, Kyoto, Japan) was developed as the first BP monitor with electrocardiogram (ECG) capability in a single device to simultaneously monitor ECG and BP readings. This study investigated whether the Complete can accurately differentiate sinus rhythm (SR) from AF during BP measurement. Fifty-six consecutive patients with persistent AF admitted for catheter ablation were enrolled in the study (mean age 65.8 years; 83.9% male). In all patients, 12-lead ECGs and simultaneous Complete recordings were acquired before and after ablation. The Complete interpretations were compared with physician-reviewed ECGs, whereas Complete recordings were reviewed by cardiologists in a blinded manner and compared with ECG interpretations. Sensitivity, specificity, and κ coefficient were also determined. In all, 164 Complete and ECG recordings were simultaneously acquired from the 56 patients. After excluding unclassified recordings, the Complete automated algorithm performed well, with 100% sensitivity, 86% specificity, and a κ coefficient of 0.87 compared with physician-interpreted ECGs. Physician-interpreted Complete recordings performed well, with 99% sensitivity, 85% specificity, and a κ coefficient of 0.85 compared with physician-interpreted ECGs. The Complete, which combines BP and ECG monitoring, can accurately differentiate SR from AF during BP measurement.
More Related Videos
08:10Estimating Bilateral Atrial Function by Cardiovascular Magnetic Resonance Feature Tracking in Patients with Paroxysmal Atrial Fibrillation
Published on: July 20, 2022
16:40A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Related Concept Videos
Dysrhythmias V: Evaluating Dysrhythmias
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Measurement of Blood Pressure
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure: