Related Experiment Video
Updated: Nov 13, 2025

Translational Rabbit Model of Chronic Cardiac Pacing
Published on: January 6, 2023
Major adverse clinical events associated with implantation of a leadless intracardiac pacemaker
Robert G Hauser1, Charles C Gornick1, Raed H Abdelhadi1
1Minneapolis Heart Institute Foundation, Minneapolis, Minnesota.
Background:
Leadless intracardiac pacemakers were developed to avoid the complications of transvenous pacing systems. The Medtronic Micra™ transcatheter pacemaker is one such system. We found an unexpected number of major adverse clinical events (MACE) in the Food and Drug Administration's Manufacturers and User Facility Device Experience (MAUDE) database associated with Micra implantation.
Objective:
The purpose of this study was to describe these MACE and compare them to implant procedure MACE in MAUDE for Medtronic CapSureFix™ active-fixation transvenous pacing leads.
Methods:
During January 2021, we queried the MAUDE database for reports of MACE for Micra pacemakers and CapSureFix leads using the simple search terms "death," "tamponade," and "perforation." Reports from 2016-2020 were included.
Results:
The search identified 363 MACE for Micra and 960 MACE for CapSureFix leads, including 96 Micra deaths (26.4%) vs 23 CapSureFix deaths (2.4%) (P <.001); 287 Micra tamponades (79.1%) vs 225 tamponades for CapSureFix (23.4%) (P <.001); and 99 rescue thoracotomies for Micra (27.3%) vs 50 rescue thoracotomies for CapSureFix (5.2%) (P <.001). More Micra patients required cardiopulmonary resuscitation (21.8% vs 1.1%) and suffered hypotension or shock (22.0% vs 5.8%) than CapSureFix recipients (P <.001). Micra patients were more likely to survive a myocardial perforation or tear if they had surgical repair (P = .014).
Conclusion:
Micra leadless pacemaker implantation may be complicated by myocardial and vascular perforations and tears that result in cardiac tamponade and death. We estimate the incidence is low (<1%). Rescue surgery to repair perforations may be lifesaving. MACE are significantly less for implantation of CapSureFix transvenous ventricular pacing leads.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
16:40A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Related Concept Videos
Mechanism of Cardiac Arrhythmias
Cardiomyopathy V: Interprofessional Care
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Mitral Valve Prolapse II: Assessment and Management
Introduction Cardiac Emergencies
Cardiac Catheterization II: Right Heart Catheterization