Related Experiment Video
Updated: Nov 13, 2025

Portable Thermographic Screening for Detection of Acute Wallenberg's Syndrome
Published on: September 19, 2019
Globally deployed COVID-19 fever screening devices using infrared thermographs consistently normalize high readings
Conor Healy1, Zachary Segal1, Chris Hinnerichs2
1IPVM, Bethlehem, Pennsylvania, United States.
Significance:
The need for regulatory review of infrared thermographs (IRTs) used on humans was removed in response to the unique circumstances of the SARS-CoV-2 pandemic (a.k.a., COVID-19). The market for these devices has since expanded considerably. This evaluation of IRT performance may have significant implications for febrility screening worldwide.
Aim:
Perform controlled nonhuman trials of IRT devices to identify and quantify deviations in the human temperature range.
Approach:
We compared IRT readings of a temperature-controlled non-human subject with one FDA-cleared IRT and one FDA-cleared handheld NCIT. In individual trials for each device, the subject was measured between 35°C and 40°C at 0.25°C increments.
Results:
The IRT device measurements were consistently normalized around the human mean (∼37 ° C). Temperatures were decremented as they approached the febrile range, and systematically reported as normal across all seven devices. This effect does not appear to be explained by a fixed offset or any known approach to estimating body temperature, or by random error.
Conclusion:
The IRTs in this study generated human temperature measurements that were systematically biased to the mean human temperature. Given that these devices are utilized for sentinel detection of possible infectious disease transmission, and are now globally employed, the implications for reduced detection of febrility are a widespread false sense of security. This vulnerability must be considered with respect to facility access control, clinical applications, and travel screening in the context of the ongoing COVID-19 pandemic response.
Related Concept Videos
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Increased Body Temperature