Analysis of bone architecture using fractal-based TX-Analyzer™ in adult patients with osteogenesis imperfecta
Jakob E Schanda1, Stephanie Huber2, Martina Behanova2
1AUVA Trauma Center Vienna-Meidling, Department for Trauma Surgery, Vienna, Austria; Ludwig Boltzmann Institute for Experimental and Clinical Traumatology, Vienna, Austria; Austrian Cluster for Tissue Regeneration, Vienna, Austria.
Background:
Osteogenesis imperfecta (OI) is a rare genetic disorder characterized by impaired bone quality and quantity. Established imaging techniques have limited reliability in OI. The TX-Analyzer™ is a new, fractal-based software allowing a non-invasive assessment of bone structure based on conventional radiographs. We explored whether the TX-Analyzer™ can discriminate OI patients and healthy controls. Furthermore, we investigated the correlation between TX-Analyzer™ parameters and (i) bone mineral density (BMD) by Dual Energy X-ray Absorptiometry (DXA), (ii) trabecular bone score (TBS), and (iii) bone microstructure by high-resolution peripheral quantitative computed tomography (HR-pQCT).
Material And Methods:
Data of 29 adult OI patients were retrospectively analyzed. Standard radiographs of the thoracic and lumbar spine were evaluated using the TX-Analyzer™. Bone Structure Value (BSV), Bone Variance Value (BVV), and Bone Entropy Value (BEV) were measured at the vertebral bodies T7 to L5. Data were compared to a healthy, age- and gender-matched control group (n = 58). BMD by DXA, TBS, and trabecular bone microstructure by means of HR-pQCT were correlated to TX-Analyzer™ parameters in OI patients. The accuracy of the TX-Analyzer™ parameters in detecting OI was assessed with area under curve (AUC) analysis of receiver operating characteristic (ROC).
Results:
BEV of the thoracic and the lumbar spine were significantly lower in OI patients compared to controls (both p < 0.001). BEV of the thoracic spine was significantly correlated to TBS (ρ = 0.427, p = 0.042) as well as trabecular number (Tb.N) at the radius (ρ = 0.603, p = 0.029) and inhomogeneity of the trabecular network (Tb.1/N.SD) at the radius (ρ = -0.610, p = 0.027), when assessed by HR-pQCT. No correlations were found between BEV and BMD by DXA. BEV of the thoracic and the lumbar spine had an AUC of 0.81 (95% confidence interval [CI] 0.67-0.94, p < 0.001) and 0.73 (95% CI 0.56-0.89, p = 0.008), respectively. BSV and BVV did not differ between OI patients and controls.
Conclusion:
The software TX-Analyzer™ is able to discriminate patients with OI from healthy controls. ROC curves of BEV values suggest a suitable clinical applicability. Low to no correlations with conventional methods suggest, that the TX-Analyzer™ may indicate a new and independent examination tool in OI.
More Related Videos
06:59Author Spotlight: An Economic and Efficient Method for Quantitative Evaluation of Bone Microarchitecture in a Murine Osteoporosis Model
Published on: September 8, 2023
08:39Longitudinal Micro-Computed Tomography Image Analysis for User-Defined Region of Interest in Critical-Sized Bone Defects
Published on: June 24, 2025
Related Concept Videos
Bone Remodeling
Fractures: Bone Repair
Minor fractures with no bone displacement are treated by immobilizing the fractured bone using a cast or splint. However, in the case of fractures with displaced bones, the broken bones are repositioned before immobilization to ensure successful healing without deformation and loss of function. The realignment of fractured bone ends is performed through a process called reduction. If the...
Bone Disorders
Bone deposition is also affected by the levels of sex hormones like estrogen and testosterone that promote osteoblast activity and bone matrix synthesis. When the level of these hormones decreases due to aging, it causes a reduction in bone deposition. As a result, bone resorption by osteoclasts...